Abstract
Background
The management of bleeding is paramount to any surgical procedure. With the increased use of less invasive laparoscopic and robotic methods, achieving hemostasis can be challenging since the surgeons cannot manually apply hemostatic agents directly onto bleeding tissue. In this study, we assessed the use of a pliable hemostatic sealant patch comprising fibrous gelatin carrier impregnated with poly(2-oxazoline) (NHS-POx) for hemostasis in robotic liver resection in a porcine bleeding model.
Methods
The NHS-POx-loaded patch (GATT-Patch), was first evaluated in a Feasibility Study to treat surgical bleeding in 10 lesions, followed by a Comparative Study in which the NHS-POx patch was compared to a standard-of-care fibrin sealant patch (TachoSil), in 36 lesions (superficial, resection, or deep injuries mimicking metastasectomies). For each lesion type, the NHS-POx and fibrin sealant patches were used in an alternating fashion with 18 lesions treated with NHS-POx and 18 with the fibrin patch. Animal preparation and surgical procedures were consistent across studies. The primary outcome was time to hemostasis (TTH) within 3min for the Feasibility Study and within 5min for the Comparative Study.
Results
In the Feasibility Study, 8 of the 10 NHS-POx-treated lesions achieved hemostasis at 30s and 3min. In the Comparative Study, all 18 NHS-POx patch-treated lesions and 9 of the 18 fibrin sealant patch-treated lesions achieved hemostasis at 5min. Median TTH with NHS-POx vs fibrin sealant patch was 30 vs 300s (P < 0.001).
Conclusions
In this animal study, hemostasis during robotic liver surgery was achieved faster and more often with the NHS-POx loaded vs fibrin sealant patch.
Supplementary Information
The online version contains supplementary material available at 10.1186/s12893-023-02159-4.
Keywords: Bleeding, Hemostasis, NHS-POx patch, GATT-Patch, TachoSil, Fibrin sealant patch, Liver surgery, Robotic
Background
For many surgical indications open procedures are being replaced by less invasive laparoscopic and robotic methods. Minimally invasive liver surgery is internationally accepted as a safe and feasible alternative to open hepatectomy. It is associated with lower perioperative morbidity, less blood loss and shorter hospitalization time, while maintaining equivalent oncological outcomes [1–4].
During any type of surgery there may be surgical bleeding for which hemostasis is difficult to achieve, necessitating the use of topical hemostatic agents [5–10]. However, the risk of substantial bleeding during minimally invasive liver surgery is a major concern. The magnitude of intraoperative blood loss is independently associated with worse perioperative and long-term outcomes [11–13]. Control of bleeding with topical hemostatic agents may be particularly challenging in these minimally invasive procedures since surgeons cannot manually apply hemostatic agents directly to the bleeding site. This could result in prolonged bleeding, requirement for blood transfusions, and potentially conversion from minimally invasive to open surgery.
In order for a topical hemostatic agent to be effective in minimally invasive surgery, the agent should be flexible and easy to handle for introduction through a trocar, successful navigation towards a bleeding site with subsequent adequate positioning, and be able to handle at least moderate bleeding severity to prevent the need for conversion from minimally invasive to open surgery. However, currently available topical hemostatic agents have a number of limitations, such as not being approved for minimally invasive use, not user-friendly to achieve successful introduction, navigation, and positioning, and not sufficiently effective for moderate or severe bleedings [14–16]. During minimally invasive surgeries, hemostatic powders or flowable matrix-based absorbable hemostats are often preferred, but these generally lack the ability to stop problematic bleedings [17].
GATT-Patch (GATT Technologies BV, Nijmegen, The Netherlands) is a new poly(2-oxazoline) (NHS-POx)-based hemostatic sealant patch that has specific design features for hemostatic performance even in severe bleeding and with flexibility and pliability to allow ease of use in minimally invasive procedures [18, 19]. In previous animal studies, use of this NHS-POx-loaded patch has demonstrated safety and efficacy during open procedures on the liver and spleen, and during robotic procedures on the kidney [19, 20].
In order to evaluate the efficacy of the NHS-POx loaded patch in minimally invasive surgery, we performed robotic liver resection in a porcine bleeding model compared to the standard-of-care fibrin sealant patch (TachoSil), for the achievement of hemostasis.
Methods
Study designs
Initial use of the NHS-POx-loaded patch was evaluated in a Feasibility Study in a porcine model with 2 animals. The primary objective of the Feasibility Study was to familiarize a single surgeon experienced with robotic procedures with NHS-POx patches, who regularly uses the fibrin sealant patch, i.e. control arm, during robotic liver surgery in clinical practice. A specific surgical procedure was not prescribed for lesion creation and location of bleeding sites, so that the surgeon would be trained on the NHS-POx patch in a variety of scenarios without limitations.
Subsequent to the Feasibility Study, the NHS-POx-loaded patch was compared to the fibrin sealant patch in 4 animals in which 36 lesions were attempted (Comparative Study), which included specific procedures for lesion creation and location of bleeding sites. For each lesion type, the NHS-POx and fibrin sealant patches were used in an alternating fashion with 18 lesions treated with NHS-POx and 18 with the fibrin patch. No a priori sample size calculation was performed.
The same methods were used in both studies for the preparation and treatment of the animals. The studies were approved by the Ethics Committee and performed in the ORSI facility at Proefhoevestraat 18, 9090 Melle, Belgium. All ORSI animal programs are in compliance with Belgian and European legislation on protection of animals used for scientific purposes.
Study devices
We examined the efficacy of an NHS-POx-loaded patch to achieve hemostasis in minimally invasive robotic porcine liver bleeding model. For comparison, a fibrin sealant patch (TachoSil Patch(Corza Medical, Westwood, Massachusetts, USA) was used, which is approved in the European Union for minimally invasive use as a method to achieve hemostasis during surgical procedures [14].
Study animals
Six (5 male, 1 female) Rattlerow Seghers (Suffolk, UK) pigs, age 13–15weeks and weighing 50–60kg, were used during these terminal experiments. The animals were sedated using ketamine (10mg/kg), medetomidine (40µg/kg) and morphine (0.1mg/kg), and subsequently anesthetized with propofol (1–3mg/kg). Intubation was performed after 10% lidocaine spray application on the trachea. Anesthesia was maintained during the procedure using isoflurane/sevoflurane in 55% oxygen. Intraoperatively the animals further received Ringer’s lactate solution (10ml/kg/h). A mean arterial blood pressure higher than 60mm Hg was maintained using L-noradrenaline, if required. Animals were monitored by and for electrocardiography, pulse oximetry, capnography and spirometry, invasive blood pressure, and temperature. At the end of the experiment, animals were euthanized with a lethal dose of pentobarbital (50mg/kg) or T61 (1ml/10kg).
Procedures
Animals were placed in supine position on the operating table in front of a Da Vinci Robot (Intuitive, Sunnyvale, California, USA). After anaesthesia, using a Veress needle and CO2 influx, 3 standard robotic 8-mm ports were introduced across the abdomen for the robotic arms, 1 standard 12-mm port for the camera, and 1 (in the Comparative Study) or 2 (in the Feasibility Study) standard 12-mm ports were placed for suction, introduction of sutures if required, and introduction of gauzes and hemostatic patches (NHS-POx loaded or fibrin sealant). The intra-abdominal pressure was set at 10mm Hg in the Feasibility Study and 8mm Hg in the Comparative Study during the entire procedure.
Sharp non-cauterizing scissors, sharp monopolar-cauterizing scissors, or bipolar vessel sealing systems were used to create lesions of the liver parenchyma, and the main vessels of the liver in the Feasibility Study (Fig.1). The goal was to create 4–6 lesions per animal in the Feasibility Study and 8 10 lesions per animal in the Comparative Study, depending on the size, anatomy and position of the liver, the location of the lesions, and the hemodynamic status of the animal.
Fig. 1.
After lesion creation, the severity of bleeding was adjudicated according to the Severity Bleeding Surface Scale (SBSS) by consensus of 2 trained investigators (E.A.R. and S.J.H.) that hold a certificate on the adjudication of bleeding on this scale of 0–5 bleeding severity: 0 represents no bleeding, 1 represents minimal bleeding, 2 represents mild bleeding, 3 represents moderate bleeding, 4 represents severe bleeding, and 5 represent extreme, life-threatening bleeding [21]. Hemostasis was considered to be an SBSS of 0.
NHS-POx-loaded and fibrin sealant patches were cut to the appropriate size so that it overlapped non-bleeding tissue by at least 1cm on all sides. The NHS-POx patch was subsequently rolled over its long axis and grasped at the most distal tip to be introduced through a 12-mm size trocar. After navigation to the bleeding and unrolling, the NHS-POx patch was positioned on the bleeding site and pressure applied for 30s with a saline-wetted gauze per Instructions for Use (IFU), after which hemostasis was checked by carefully removing the gauze (Fig.1). If hemostasis was not yet achieved, either additional pressure, or additional (pieces of) NHS-POx patch was applied. Hemostasis was checked at predetermined times of 30s, 1min, and 3min in both the Feasibility and Comparative Study, with the additional primary endpoint of 5min only for the Comparative Study. The time to hemostasis (TTH) was considered to be the time that hemostasis was achieved and maintained until the end of the procedure, which was up to about 1 three-fourths hours after creation of the first lesion.
In the Feasibility Study, the NHS-POx-loaded patch was primarily used to train the surgeon, but the fibrin sealant patch was used as a control on 1 lesion for direct comparison of ease of use in a single surgery. For this lesion, the fibrin sealant patch was folded and placed inside a cut-off finger from a surgical glove, to improve ease of introduction through the trocar, after which it was grasped at the most distal tip and introduced through a 12- mm trocar. In the Comparative Study, the fibrin sealant patch was folded in a zigzag shape over its long axis and introduced directly without placing it in the surgical glove. After introduction, the patch was navigated to, and positioned on the bleeding site and pressure was applied for 3min (per IFU) with a saline-wetted gauze, after which the gauze was carefully removed and hemostasis was checked. If hemostasis was not yet achieved, either additional pressure, or additional (pieces of) fibrin sealant patch was applied up to the time of the primary endpoint of 5min when hemostasis was always checked.
Outcomes
The primary endpoint for both studies was hemostatic performance as determined by the TTH, with a pass or fail at 3min in the Feasibility Study and at 5min for the Comparative Study. Since the fibrin sealant patch has a prescribed 3-min application time when hemostasis is checked, we extended the TTH outcome to 5min in the Comparative Study to confirm complete hemostasis was achieved with no rebleeding. Additional secondary outcomes included the use of the NHS-POx-loaded patch and fibrin sealant patch with a measure of pass (successful procedure without damaging the patch) or fail (patch could no longer be used). These measures included preparation and introduction of the patch through a trocar, navigation to the bleeding site, unrolling or unfolding of the patch, positioning on the bleeding site, application of the patch with a wet gauze and the removal of the gauze after placement.
Statistical analyses
Results are provided as descriptives. The difference in median TTH between groups was calculated with the Mann–Whitney U test, with a P value of < 0.05 defining statistical significance. When hemostasis was not achieved and a TTH was not available, a time of 305s (5s longer than the primary endpoint of hemostasis at 5min) was used for input in the calculation.
Results
Feasibility study
A total of 8 liver lesions were created (4 resections, 4 mimicking metastasectomies). The NHS-POx-loaded patch was used in 7 lesions and the fibrin sealant patch was attempted in 1 lesion. The median interquartile range (IQR) SBSS was 4 (3–4). The NHS-POx patch achieved hemostasis in 30s in 6 of 7 (86%) lesions (Table 1). One lesion, with extreme, life-threatening bleeding (SBSS 5), required an additional NHS-POx patch placement after a piece of patch came loose and hemostasis was achieved at 8min and did not meet the primary endpoint.
Table 1.
Feasibility study lesion details
Lesion | Lesion characteristics and size | Patch | Patch Size | SBSS | TTH (s) | Remarks |
---|---|---|---|---|---|---|
1 | Metastasectomy 1 × 1cm and 0.8cm depth | NHS-POx | 1 × 2cm in depth of the lesion and 4 × 5cm on top | 3 | 30 | None |
2 | Resection 5 × 1.5cm and 2cm depth | NHS-POx | 6 × 5cm | 4 | 30 | None |
3 | Resection 6 × 2cm and 3cm depth | NHS-POx | 1 full patch (10 × 5cm) | 3 | 30 | None |
4 | Metastasectomy 3.5 by 3cm and 2cm depth | NHS-POx | 1 full patch (10 × 5cm) | 2 | 30 | Piece of patch came loose upon removal of gauze, but hemostasis achieved |
5 | Metastasectomy 2.5 by 2.5cm and 2cm depth | NHS-POx | 1 full patch (10 × 5cm) in depth of the lesion and 5 × 5cm on top | 4 | 30 | None |
6 | Portal vein – 0.25 × 0.25cm | NHS-POx | 1/3 patch | 4 | 30 | None |
7 | Metastasectomy 1.5 by 1.5cm and 1.5cm depth | NHS-POx | 3.5 × 5cm in depth of the lesion and 3.5 × 5cm on top | 4 | 30 | None |
8 | Resection 3 by 2cm and 3cm depth | NHS-POx | 6.5 × 5cm and 5 × 5cm | 5 | 480 | Piece of patch came loose, additional patch applied resulting in hemostasis |
9 | Resection 3 by 2cm and 3cm depth | Fibrin sealant | 4.8 × 8cm and 4.8 × 4.8cm | 4 | Hemostasis not achieved | First patch broke in trocar, no hemostasis achieved |
9 | Resection 3 by 2cm and 3cm depth | NHS-POx | 1 full patch (10 × 5cm) | 4 | 30 | Used after TachoSil failed |
10 | Vena cava | NHS-POx | NA | 5 | Hemostasis not achieved | First attempt failed; second attempt achieved hemostasis. After removal of gauzes and manipulation of bleeding site, rebleeding occurred and hemostasis was not achieved with 2 additional attempts |
Open in a new tab
NA not available, SBSS Severity Bleeding Surface Scale, TTH time to hemostasis
For the 1 lesion treated with the fibrin sealant patch, it was introduced through the trocar before creating the bleeding lesion. During introduction, a small part of the patch broke off and remained in the trocar. After the lesion was created (SBSS 4), and as the fibrin sealant patch was removed from the glove, additional pieces broke off, but could still be placed on the bleeding site. After 2 attempts of holding pressure with a wet gauze for 3min, hemostasis was achieved, not meeting the primary endpoint. Subsequently, breakthrough bleeding occurred and eventually the fibrin sealant patch was removed and a NHS-POx-loaded patch was placed with hemostasis achieved in 30s.
After the liver resection procedures, 1 vascular lesion was created in each animal in a large vein (portal vein and inferior vena cava) to attempt to treat severe and extreme, life-threatening bleeding. The cut in the portal vein (SBSS 4) was successfully treated with the NHS-POx-loaded patch with hemostasis achieved at 30s. The cut in the vena cava (SBSS 5) was large (about 5mm) and could only be visualized with several robotic arms, which required removing tools that would otherwise be used to perform suction, grasp gauzes, or introduce the patch. Hemostasis was initially achieved, but after removal of gauzes with manipulation of the site, rebleeding occurred. With compromised visualization and after multiple attempts, persistent hemostasis could not be achieved.
Comparative study
A total of 36 lesions were created and treated with NHS-POx-loaded patch (n = 18) or fibrin sealant patch (n = 18). Of these, 16 were superficial lesions (n = 8 for both arms), 8 (n = 4 for both arms) were partial liver resections (see Fig.1 for example), and 12 (n = 6 for both arms) were deep mimicking metastasectomies (see Fig.2 for example). For the superficial lesions, the median SBSS was 2 in each group, for the resections, the median SBSS was 3 in each group, and for mimicking the deep metastasectomies, the median SBSS was 4 in each group (Table 2). The median (interquartile range [IQR]) SBSS for all lesions was 3 (2–4) for both groups.
Fig. 2.
Open in a new tab
Table 2.
Comparative study lesion details
NHS-POx Patch N = 18 | Fibrin Sealant Patch N = 18 | |||
---|---|---|---|---|
SBSS | Time to hemostasis (s) | SBSS | Time to hemostasis (s) | |
Superficial | 2 | 30 | 2 | Not achieved |
3 | 150 | 3 | Not achieved | |
2 | 30 | 2 | 180 | |
2 | 30 | 2 | 180 | |
2 | 30 | 2 | 180 | |
1 | 30 | 2 | 300 | |
2 | 30 | 1 | 180 | |
2 | 30 | 2 | Not achieved | |
Resection | 3 | 280 | 3 | Not achieved |
2 | 30 | 2 | Not achieved | |
3 | 30 | 4 | Not achieved | |
4 | 30 | 3 | 180 | |
Deep metastasectomy | 4 | 280 | 3 | Not achieved |
5 | 30 | 5 | 180 | |
4 | 30 | 5 | 300 | |
4 | 30 | 4 | Not achieved | |
5 | 30 | 4 | 300 | |
3 | 30 | 3 | Not achieved | |
All lesions, median (IQR) | 3 (2–4) | 30 (30–30) | 3 (2–4) | 300 (180 – not achieved) |
Open in a new tab
IQR interquartile range, SBSS Severity Bleeding Surface Scale
Hemostasis was achieved at the primary endpoint at 5min in all 18 (100%) NHS-POx-loaded patch lesions and 9 of the 18 (50%) of the fibrin sealant patch lesions (Fig.3). NHS-POx patch achieved hemostasis in 15 of 18 lesions (83%) after the IFU-indicated 30s pressure and the fibrin sealant patch achieved hemostasis in 6 of 18 (33%) lesions after the IFU-indicated 3min of pressure. The overall median (IQR) TTH was 30s (30s – 30s) in the NHS-POx patch treated lesions and 300s (180s – not achieved) for the fibrin sealant lesions, with a P value of < 0.001 (Table 2). The mean TTH was shorter with NHS-POx patch for all lesion types.
Fig. 3.
Open in a new tab
Usability
For all 28 liver and vascular lesions treated with the NHS-POx-loaded patch in both studies, 42 pieces of or full-sized NHS-POx patch were introduced. The introduction through the trocar was successful without issues in all but 2 cases: (1) when used not in accordance with IFU, there was a small tear of a piece of NHS-POx patch that was folded over its short axis instead of rolled over its long axis (per IFU) for introduction, making it thicker, though it could still be used for successful hemostasis; (2) NHS-POx patch was temporarily adhering within the trocar due to excessive wetness of the trocar, but could be delivered and used successfully for hemostasis with a minor delay of less than 30s. Subsequent navigation to the lesion, unfolding the patch and positioning on the bleeding site was successful without adhesion of the patch to other tissue, tearing, or difficulty to position.
For all 19 lesions attempted with the fibrin sealant patch, and used in accordance with IFU, the patch could be introduced through the trocar, and it was possible to navigate to the lesion and unfold the patch and position on bleeding site. Yet in 6 cases, the surgeon reported more difficulty to unfold the fibrin sealant patch as compared to NHS-POx patch. In all 19 cases, it was possible to position the patch on the bleeding site and loosen the gauze from the patch, although with removal of the gauze the patch frequently loosened from the tissue at the edges.
Discussion
There is an unmet need of surgical tools to treat bleeding during minimally invasive liver surgery to further optimize clinical outcomes [11–13]. In these first studies of an NHS-POx-loaded patch in minimally invasive robotic liver resections in a porcine bleeding model, hemostasis of liver bleeding, including severe bleeding, can be achieved quickly and reliably. It could be introduced through the 12-mm trocar in all lesions, could be easily navigated to and be positioned onto the bleeding site with adequate pressure upon application. When compared to a standard-of-care hemostatic patch, hemostasis at 5 min was achieved in 25 of the 26 (96%) NHS-POx patch lesions as compared to 9 out of 19 (47%) fibrin sealant patch lesions. Moreover, the median TTH was significantly shorter with the NHS-POx patch vs the fibrin sealant patch (30 vs 300 s; P < 0.001).
Several important differences exist between the NHS-POx-loaded patch and the fibrin sealant patch. First, the fibrin sealant patch is a biologically active patch with thrombin and fibrinogen to induce blood clotting, while the NHS-POx patch does not include any biologically active components; it consists of a synthetic polymer impregnated in a gelatin carrier, which allows the patch to adhere strongly to the bleeding tissue and provides a hemostatic seal at the site of application. Second, the carrier for the fibrin sealant patch is an equine collagen, which is brittle and not easily handled during minimally invasive procedures in this study. In fact, in 3 lesions, pieces of the active components of the fibrin sealant patch broke off in the trocar or while placing the patch. On the contrary, the carrier of the NHS-POx patch is a fibrous structure of porcine gelatin which makes it flexible and pliable. In our study, all NHS-POx and fibrin sealant patches could be delivered through the trocar and navigated to the lesion and positioned on the bleeding site. Despite the surgeon being experienced with the fibrin sealant patch, it was noted that the patch was difficult to deliver in some lesions and would loosen at the edges with removal of the gauze. Third, the NHS-POx patch does not require additional preparation after rolling to introduce into the trocar, compared with careful folding and sometimes introduction in a cut-off finger from a surgical glove for the fibrin sealant patch.
These design aspects result in a number of clinical differences, including the time required by the surgeon to apply pressure after placement. The NHS-POx-loaded patch has an application period of 30s to achieve hemostasis, versus 3min for the fibrin sealant patch, and in our studies, 26 NHS-POx patch liver applications were performed with 100% success (85% achieved hemostasis in 30s). While in the lesions treated with the fibrin sealant patch, only 30% achieved hemostasis at 3min (per IFU), and only 50% reached hemostasis at the primary endpoint of 5min.
The NHS-POx-loaded patch achieved TTH of 30s consistently across all 3 lesion types (Table 2), while the fibrin sealant patch often required additional pressure holding if hemostasis was not achieved. This difference in TTH, especially for the resection in a thin porcine liver, demonstrates the flexibility and ease-of-use of the NHS-POx-loaded patch to fold around the bleeding lesion and position on complex bleeding in minimally invasive procedures. In deep injuries mimicking metastasectomies, with the NHS-POx patch, 5 of the 6 achieved a TTH of 30s, and 1 of the 6 achieved a TTH of 3min with the fibrin sealant patch. In deep lesions, the fibrous NHS-POx patch can be placed into the depth of the lesion to "plug" the hole, and additional pieces can be added, if necessary. This is not possible with the brittle traditional flat surfaced fibrin sealant patch, thus is a unique feature of the NHS-POx patch that provides an advantage even over flowable matrix-based absorbable hemostats that are often used in these types of bleeding [22].
In this study we demonstrated that the NHS-POx-loaded patch may not be limited to mild to moderate bleeding like other hemostatic agents, but can also be successfully used in severe bleeding. Severe bleeding remains a challenge in liver surgery, especially using a minimally invasive laparoscopic approach [23]. Few products are currently available in clinical practice and are not as effective in severe bleeding. In our studies, there were 17 severe or extreme, life-threatening bleeding (SBSS 4–5) liver resection and lesions mimicking deep metastasectomies (11 in the NHS-POx patch group [including the rescue] and 6 in the fibrin sealant patch group). One bleeding was first attempted with the fibrin sealant patch, but required rescue treatment with the NHS-POx patch. Of the 11 NHS-POx patch lesions, 9 (82%) achieved hemostasis in 30s; of the 6 fibrin sealant patch lesions, 1 achieved hemostasis in 3min. Our studies included 2 vascular lesions for which the use of the NHS-POx patch was considered experimental in nature for potential future studies. Nevertheless, in the portal vein lesion with SBSS 4, the NHS-POx patch achieved hemostasis in 30s. In the vena cava lesion with SBSS 5, the NHS-POx patch could not be placed accurately on the bleeding site due to the difficulty in visualizing the bleeding site. The use of the NHS-POx patch on vascular lesions was not the main objective of this study, however, future studies are warranted as a primary method for achieving hemostasis or as an adjunct in combination with suturing.
Our study was performed in a porcine model for the initial use of a new hemostatic patch during minimally invasive robotic liver surgery. As the NHS-POx patch does not currently have regulatory approval for clinical use (i.e., conformité européenne [CE] Mark or United States Food and Drug Administration [FDA] approval), an evaluation of minimally invasive use was limited to an animal model. For the evaluation of a hemostatic agent, continuous blood flow is required and although ex-vivo perfusion of selective organs provides an alternative to live animal surgery, the translational value is limited, specifically in the setting of minimally invasive surgery. Furthermore, we found the model to be suitable for future training with the product before clinical application after product approval. Specifically, because other training models, such as simulation, have not advanced to the point in which the operator’s experience includes managing real-life pathophysiology. This animal model provides a unique learning experience and was effective to allow the operator to perform procedures on a variety lesion types with a variety of bleeding severity. The technical skills gained from this study can be used to support robotic liver surgery procedures in humans.
Conclusions
The data from this preclinical study provide evidence that hemostasis of up to life-threatening bleeding during robotic liver surgery can be achieved quicker and more reliably with the NHS-POx-loaded patch when compared to the standard of care product, a fibrin sealant patch. In addition to the hemostatic performance, the lack of preparation time and the ease-of-use of the NHS-POx patch, even in minimally invasive settings, may help reduce the need for conversion to open surgery to control bleedings and improve clinical outcomes. Future studies are needed to evaluate the use of GATT-Patch in clinical procedures.
Supplementary Information
Additional file 1. (87.4KB, pdf)
Acknowledgements
Jane Moore, MS, ELS, MSRM, llc, drafted the manuscript in collaboration with S.J.H. and created the graphical figure and tables funded by GATT Technologies.
Authors’ contributions
M.D., E.A.R., J.B. and S.J.H. designed the study protocol; M.D. and F.N. performed the surgical procedures; E.A.R., K.B. and S.J.H. collected the procedural data and supported the animal procedures; A.M. supported the animal procedures and represented the ORSI animal facility; J.B. and S.J.H. served as Sponsor representatives. E.A.R. prepared all surgical figures. All authors critically reviewed the manuscript and approved the final version.
Funding
The studies were sponsored by GATT Technologies B.V., located at Novio Tech Campus on Transistorweg 5, 6534 AT Nijmegen, The Netherlands.
Availability of data and materials
All data generated or analyzed during this study are included in this published article.
Declarations
Ethics approval and consent to participate
The Ethical Committee (EC number 2021/072) of the Faculty of Veterinary Medicine and the Faculty of Bioscience Engineering of Ghent University approved all experimental protocols reported in this manuscript. All methods were carried out in accordance with relevant guidelines and regulations. All study methods were performed in accordance with the ARRIVE guidelines for the reporting of animal experiments.
Consent for publication
Not applicable.
Competing interests
Prof. D’Hondt has received remuneration for preclinical services from GATT Technologies. Dr. Nuytens has received consulting fees for preclinical testing of the product from GATT Technologies. Mr. Roozen provided study materials and assisted in the writing of this manuscript and is an employee of GATT Technologies. Dr. Bender provided study materials and assisted in the writing of the manuscript and was an employee and shareholder of GATT Technologies through May 2022. Dr. Bender is a coinventor on patent WO 2021/009013. Dr. Head provided study materials, analyses of the data, medical writing and any article processing charges, if applicable, and was an employee and shareholder of GATT Technologies through May 2022. Dr. Head has a patent pending. Dr. Head and Dr. Bender are currently consultants of GATT Technologies. Dr. Mottrie has received consulting fees and honoraria from Medtronic and is a shareholder of ORSI Academy. Mr. Bauwens is an employee of ORSI Academy.
Footnotes
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Croome KP, Yamashita MH. Laparoscopic vs open hepatic resection for benign and malignant tumors: an updated meta-analysis. Arch Surg. 2010;145:1109–1118. doi: 10.1001/archsurg.2010.227. [DOI] [PubMed] [Google Scholar]
- 2.Hildebrand N, Verkoulen K, Dewulf M, Heise D, Ulmer F, Coolsen M. Short-term outcomes of laparoscopic versus open hepatectomy in the elderly patient: systematic review and meta-analysis. HPB (Oxford) 2021;23:984–993. doi: 10.1016/j.hpb.2021.01.016. [DOI] [PubMed] [Google Scholar]
- 3.Kasai M, Cipriani F, Gayet B, Aldrighetti L, Ratti F, Sarmiento JM, Scatton O, Kim KH, Dagher I, Topal B, Primrose J, Nomi T, Fuks D, Abu Hilal M. Laparoscopic versus open major hepatectomy: a systematic review and meta-analysis of individual patient data. Surgery. 2018;163:985–995. doi: 10.1016/j.surg.2018.01.020. [DOI] [PubMed] [Google Scholar]
- 4.Wang ZY, Chen QL, Sun LL, He SP, Luo XF, Huang LS, Huang JH, Xiong CM, Zhong C. Laparoscopic versus open major liver resection for hepatocellular carcinoma: systematic review and meta-analysis of comparative cohort studies. BMC Cancer. 2019;19:1047. doi: 10.1186/s12885-019-6240-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Boonstra EA, Molenaar IQ, Porte RJ, de Boer MT. Topical haemostatic agents in liver surgery: do we need them? HPB (Oxford) 2009;11:306–310. doi: 10.1111/j.1477-2574.2009.00065.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Ollinger R, Mihaljevic AL, Schuhmacher C, Bektas H, Vondran F, Kleine M, Sainz-Barriga M, Weiss S, Knebel P, Pratschke J, Troisi RI. A multicentre, randomized clinical trial comparing the Veriset haemostatic patch with fibrin sealant for the management of bleeding during hepatic surgery. HPB (Oxford) 2013;15:548–558. doi: 10.1111/hpb.12009. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Moench C, Mihaljevic AL, Hermanutz V, Thasler WE, Suna K, Diener MK, Seehofer D, Mischinger HJ, Jansen-Winkeln B, Knaebel HP, Bechstein WO. Randomized controlled multicenter trial on the effectiveness of the collagen hemostat Sangustop(R) compared with a carrier-bound fibrin sealant during liver resection (ESSCALIVER study, NCT00918619) Langenbecks Arch Surg. 2014;399:725–733. doi: 10.1007/s00423-014-1203-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Verhoef C, Singla N, Moneta G, Muir W, Rijken A, Lockstadt H, De Wilt JH, Albert O, Zuckerman LA, Frohna P, Porte RJ. Fibrocaps for surgical hemostasis: two randomized, controlled phase II trials. J Surg Res. 2015;194:679–687. doi: 10.1016/j.jss.2014.12.011. [DOI] [PubMed] [Google Scholar]
- 9.Koea JB, Batiller J, Aguirre N, Shen J, Kocharian R, Bochicchio G, Garden OJ. A multicentre, prospective, randomized, controlled trial comparing EVARREST fibrin sealant patch to standard of care in controlling bleeding following elective hepatectomy: anatomic versus non-anatomic resection. HPB (Oxford) 2016;18:221–228. doi: 10.1016/j.hpb.2015.12.006. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Wells CI, Ratnayake CBB, Mentor K, Sen G, Hammond JS, French JJ, Wilson CH, Manas D, White S, Pandanaboyana S. Haemostatic efficacy of topical agents during liver resection: a network meta-analysis of randomised trials. World J Surg. 2020;44:3461–3469. doi: 10.1007/s00268-020-05621-z. [DOI] [PubMed] [Google Scholar]
- 11.Jiang W, Fang YJ, Wu XJ, Wang FL, Lu ZH, Zhang RX, Ding PR, Fan WH, Pan ZZ, De-Sen W. Intraoperative blood loss independently predicts survival and recurrence after resection of colorectal cancer liver metastasis. PLoS One. 2013;8:e76125. doi: 10.1371/journal.pone.0076125. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Margonis GA, Kim Y, Samaha M, Buettner S, Sasaki K, Gani F, Amini N, Pawlik TM. Blood loss and outcomes after resection of colorectal liver metastases. J Surg Res. 2016;202:473–480. doi: 10.1016/j.jss.2016.01.020. [DOI] [PubMed] [Google Scholar]
- 13.Hallet J, Mahar AL, Nathens AB, Tsang ME, Beyfuss KA, Lin Y, Coburn NG, Karanicolas PJ. The impact of perioperative blood transfusions on short-term outcomes following hepatectomy. Hepatobiliary Surg Nutr. 2018;7:1–10. doi: 10.21037/hbsn.2017.05.07. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Indications and Usage for TAchoSil Patch. TachoSil Patch: Package Insert / Prescribing Information - Drugs.com. Accessed 5 Dec 2022.
- 15.Ethicon Johnson & Johnson Surgical Technologies. EVARREST Fibrin Sealant Patch | Ethicon (jnjmedtech.com). Accessed 9 Nov 2022.
- 16.Medtronic Covidien. Veriset™ Hemostatic Patch | Medtronic (UK). Accessed 9 Nov 2022.
- 17.Ethicon Johnson & Johnson Surgical Technologies. SNoW Absorbable Hemostat. SURGICEL SNoW™ Absorbable Hemostat | Ethicon (jnjmedtech.com). Accessed 17 Nov 2022.
- 18.Boerman MA, Roozen E, Sanchez-Fernandez MJ, Keereweer AR, Felix Lanao RP, Bender J, Hoogenboom R, Leeuwenburgh SC, Jansen JA, Van Goor H, Van Hest JCM. Next generation hemostatic materials based on NHS-ester functionalized poly(2-oxazoline)s. Biomacromol. 2017;18:2529–2538. doi: 10.1021/acs.biomac.7b00683. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Roozen EA, Warle MC, Lomme R, Felix Lanao RP, van Goor H. New polyoxazoline loaded patches for hemostasis in experimental liver resection. J Biomed Mater Res B Appl Biomater. 2022;110:597–605. doi: 10.1002/jbm.b.34938. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Farinha R, De Groote R, Zondervan PJ, Paciotti M, Roozen E, Head S, Sarchi L, Bravi C, Mottaran A, Puliatti S, De Backer P, De Naeyer G, Mottrie A. Will renorrhaphy become obsolete? Evaluation of a new hemostatic sealant. J Endourol. 2022. 101089/end20220339. [DOI] [PubMed]
- 21.Spotnitz WD, Zielske D, Centis V, Hoffman R, Gillen DL, Wittmann C, Guyot V, Campos DM, Forest P, Pearson A, McAfee PC. The SPOT GRADE: A new method for reproducibly quantifying surgical wound bleeding. Spine (Phila Pa 1976) 2018;43:E664–e671. doi: 10.1097/BRS.0000000000002447. [DOI] [PubMed] [Google Scholar]
- 22.Roozen EA, et al. GATT-Patch hemostatic sealant patch versus flowable hemostatic matrix for severe parenchymatous organ bleeding (Publication pending).
- 23.Abu Hilal M, Underwood T, Taylor MG, Hamdan K, Elberm H, Pearce NW. Bleeding and hemostasis in laparoscopic liver surgery. Surg Endosc. 2010;24:572–577. doi: 10.1007/s00464-009-0597-x. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Additional file 1. (87.4KB, pdf)
Data Availability Statement
All data generated or analyzed during this study are included in this published article.